- Unscrew the screws on top of the box to open the PCB box lid.
- Loosen the screws in the load cell terminal (green) to disconnect the weight sensor.
- Remove the weight sensor cables from the load cell terminal.
- Detach the PCB box from the frame. The box is secured with double-sided tape. You can remove the box by moving it side to side or use a spatula.
- Remove the weight sensor cable from the box.
- Open the lid on the replacement PCB box.
- Insert the weight sensor cable into the new box.
Ensure the black rubber seal is properly in place. - Connect the weight sensor cables in the load cell terminal in the following order:
- GND: Yellow
- A+: Green
- A-: White
- GND: Black
- AVdd: Red
9. Close the lid securely.
10. Switch the BEEP base on and calibrate it in the BEEP base app.